CVE-2021-33885
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remot
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
CRITICAL · CVSS 10
EPSS 0.06921
Act now
- EPSS percentile: top 8% of all CVEs by exploitation likelihood
- Public exploit or PoC is available
- CVSS base score ≥ 7.0
Sigma rules0
YARA rules0